Strany 1 - Advanced pH/ORP Meter

1 Advanced pH/ORP Meter 850055 Instruction Manual

Strany 2 - TABLE OF CONTENTS

10 SETUP MODE displays under the “TR” icon. 3. Press MN/MX/AV. The “OUT” icon flashes on the upper display and P1.1 displays under the “OU

Strany 3

11 SETUP MODE 4. Press MN/MX/AV to enter P2.1. The default “NO” icon flashes on the middle display and P2.1 appears in the lower display. 5.

Strany 4 - INTRODUCTION

12 SETUP MODE P3.0 View Slope & Offset 1. Press MODE to select probe type as pH. 2. Press SET for 2 seconds to enter setup. 3. Press ▲ until th

Strany 5 - FEATURES

13 SETUP MODE 6. Press MN/MX/AV to enter P3.5 and view the offset val-ue. The offset value is the mV value of pH 7 and the default offset value i

Strany 6 - POWER SUPPLY

14 SETUP MODE 3. Press MN/MX/AV to enter P4.1. The default “NIS” (NIST) will flash on the LCD and P4.1 will appear on the lower portion of the

Strany 7 - LCD DISPLAY

15 SETUP MODE 3. Press MN/MX/AV to enter P5.1. “YES” will flash on the LCD display and P5.1 will appear on the lower display. 4. Press ▲ to s

Strany 8

16 SETUP MODE 5. Press MN/MX/AV to save the selection. 6. Press ESC to return to Normal Mode. P7.0 Real Time Clock Setting Thi

Strany 9

17 SETUP MODE 6. Press ESC to return to P7.0 and Normal Mode. P8.0 Reset This procedure will reset the meter to factory default settings. Me

Strany 10 - SETUP MODE

18 PH PROBE CALIBRATION Calibrate routinely before use, if the readings seem erratic, and whenever the batteries are replaced. For the highest accura

Strany 11

19 PH PROBE CALIBRATION 8. When the measured pH value is stable and the Ready function is enabled (Refer to P5.0 READY ICON page 15). “READY” will a

Strany 12

2 TABLE OF CONTENTS INTRODUCTION. . . . . . . . . . . . . . . . . . . . . . . . FEATURES . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 13

20 MEASUREMENT PROCEDURES 1. The meter is powered by 4 AAA batteries. Install batteries or connect an AC adapter to the power jack on the sid

Strany 14

21 MEASUREMENT PROCEDURES plugged into the meter. For manual temperature compensation, the default setting is 25°C. It is also poss

Strany 15

22 MEASUREMENT PROCEDURES 1. Follow Step 1 in the pH Measurement section to clean and soak the probe. 2. Press POWER to turn on. Press MODE to s

Strany 16

23 MEASUREMENT PROCEDURES Note... There is no need to take temperature compensation into consideration when using an ORP probe to measure

Strany 17

24 MEASUREMENT PROCEDURES Record Memory The meter can store up to 99 records each of pH, mV, and ORP (mV) readings. 1. In any measurement or Hold

Strany 18 - PH PROBE CALIBRATION

25 MEASUREMENT PROCEDURES Recall Maximum & Minimum This function reviews a maximum and minimum value for all the data points stored in me

Strany 19

26 MAINTENANCE pH Probe It is important to keep the pH probe wet when the meter is not in use. The probe is well protected by a plastic bottle conta

Strany 20 - MEASUREMENT PROCEDURES

27 MAINTENANCE The following actions will keep the probe in good working condition.  Always keep the pH glass bulb wet by using the plastic bott

Strany 21

28 MAINTENANCE To expose the new unused fiber portion: 1. Use tweezers to pull out the fiber junction and expose the new unused portion. 2. C

Strany 22

29 MAINTENANCE 2. Put the probe in a buffer solution of pH 7.00 with sat-urated quinhydrone. 3. Stir; mV reading (E1) should be 86 ±15mV.

Strany 23

3 TABLE OF CONTENTS Temperature Compensation. . . . . . . . . . . . . . . . Hold Function . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 24

30 MAINTENANCE When measuring solutions with a low concentration of ox-idation reduction matter and slow ion exchange rate, a slow response time

Strany 25

31 TROUBLESHOOTING Meter does not turn on: 1. Press POWER for more than 2 seconds. 2. Check connections of the power adapter or batterie

Strany 26 - MAINTENANCE

32 ERROR CODES E02 Reading is under the lower limit E03 Reading is over the upper limit E04 Preceded by another error E12 Factory calibr

Strany 27

33 PC CONNECTION The meter can interface with a personal computer to capture on-line or stored data. Connection procedures: 1. Plug the USB c

Strany 28

34 PC CONNECTION Voltage reading in mV, the third value is Temperature of pH probe in C/F, and “x” is a numerical value. Format in Memory Transmit p

Strany 29

35 APPENDIX TEMPERATURE EFFECT ON NIST pH BUFFERS

Strany 30

36 OPTIONAL ACCESSORIES 840016 pH Probe (non-ATC) 840027 AC Adapter 840049 Spear Tip pH Probe (non-ATC) 840052 Data Acquisition Software 840054 USB

Strany 31 - TROUBLESHOOTING

37 SPECIFICATIONS Operation temperature: 5°C to 40°C Operation RH%: Up to 95% w/o condensation Storage Temperature: -20°C to +60°C Storage RH%:

Strany 32 - ERROR CODES

38 SPECIFICATIONS Program Preference Default Displayed P1.0 Memory No default “tr” (Transmitting) P1.1 Memory Sent No default “out”

Strany 33 - PC CONNECTION

39 WARRANTY Sper Scientific warrants this product against defects in materials and workmanship for a period of five (5) years from the date of purch

Strany 34

4 INTRODUCTION Thank you for purchasing the Advanced pH/ORP Meter. This meter reads the pH (balance of alkalinity and acidity) and ORP (oxidation re

Strany 35 - APPENDIX

40 Advanced pH/ORP Meter 850055 Copyright ©2009 by Sper Scientific ALL RIGHTS RESERVED Printed in the USA 7720 E. Redfield Rd. Suite #7

Strany 36 - OPTIONAL ACCESSORIES

5 FEATURES  Multi-display LCD screen  Automatic buffer recognition  5 points pH calibration  Hold function  Maximum and minimum  Backl

Strany 37 - SPECIFICATIONS

6 POWER SUPPLY The meter is powered by 4 “AAA” type batteries or a 9 Volt external adapter. Insert batteries into the battery compartment

Strany 38

7 LCD DISPLAY LCD display indicates pH, mV, ORP (mV). Icon pH, mV indicates the unit of measure displayed. READY indicates the reading is stable. MAX

Strany 39 - WARRANTY

8 KEYPAD POWER/SET CAL/ESC HLD/REC MODE/▲ MEM/▼ MN/MX/AV Press to turn meter on/off. Press and hold for more than 1 second to enter SET mode. Switch

Strany 40 - Web: www.sperscientific.com

9 SETUP MODE The advanced Setup Mode allows you to customize the following meter preferences and defaults:  P1.0 Memory Transmission  P2.0 Clea