9. When “CAL” stops ashing on the
LCD, press SET for less than 1 second
to conrm the value. The pen will
return to Measurement Mode for salt
measurement. To exit Calibration Mode
without saving, press SET for more than
2 seconds. This allows you to retain the
pen’s previous calibration data for the
current range as you proceed. If “CAL”
ashes on the LCD continually, check
that the calibration solution is stable and
that the input value (Step 8) is correct.
10. Repeat Steps 1-9 for other ranges as
needed.
Note…
If the standard value is over the measuring
range or 10% less, the displayed value will
be equal to the range limit or 10% of the
range limit.