ERROR CODES
Salinity
- - - The pen is in manual ranging 1 but
the salinity measurement is higher
than 10.0 ppt; Press ▲ for more
than 2 seconds to change to manual
ranging 2 or auto-ranging mode.
E03 The salt value is over the range limit
(70.0 ppt) or the pen is damaged;
Put the pen in standard buffer
solution (the buffer concentration
must be lower than the range limit).
If E03 still appears, send the pen to
Sper Scientic for repair.
E04 Error in measuring original data
(damaged temperature sensor,
temperature out of specications
or salinity error). E04 will disappear
when the original error is resolved.